|
APDA ÀÇ Àû¿ë
- Condition
monitoring of medium voltage cable terminations and
cable joints for
in service discharge detection of defects.
- General
purpose system to detect partial discharges (PD) for other
types of equipment.
- Applicable
for both commissioning of new cable accessory and maintenance
based renewal program for existing cable system.
Cable
termiationÀÇ APDA ÀÇ Àû¿ë
- The
APDA is used to detect internal discharges.
The system has incorporated a glass fibre rod (hot stick)
that is acting as
a wave guide. The rod will transfer acoustic signals from
internal defects in
the cable accessory to the sensor.
- Location
of partial discharges is done by moving the tip of the hot
stick and
gently pressing it against the termination.
PROPERTIES
-
The instrument is monitoring the acoustic signals by an
ultrasonic sensor in an electrically insulating glass fibre
rod.
A microprocessor controls the data collection and data-processing.
-
The results are presented on a LCD screen during measurements
and may be stored and transferred to a PC via an RS-232
serial communication interface.
-
The user communicates with the instrument by means of soft
touch buttons on the front. Setting of filters, time constants
and trigger circuits are all menu controlled.
-
For easy use, recommended default settings of parameters
are included.
APDA has two measuring modes, continuous and phase mode
- The
continuous mode gives:
- RMS signal during one power cycle.
- Peak signal during one power cycle.
- Degree of modulation with the power cycle.
- Degree of modulation with twice the power cycle.
This continuous mode gives the first information about the
signal, indicating a partial discharge. Figure 1 shows the
result from measurements on a cable termination.
50 and 100 Hz content indicate that a partial discharge
is present.
- The
phase mode gives:
- Phase angle and pulse accurrance.
Figure 2 shows the correlation between the phase of power
frequency and the moment of partial discharges. It is seen
that most acoustic signals are generated around the crest
of the power frequency voltage.
SENSITIVITY
- IEC
502 specify that the partial discharge level according to
IEC 270 should be
below 10 pC.
- The
sensitivity of the APDA is in the range of 5-50 pC depending
on the design and material used in the cable accessory.
This is based on the comparison of electrically and accoustically
measured signals.
|