Continuously
Monitoring Your Environment
- Temperature
- Humidity
- Barometric
Pressure
- Line
Voltage
- Line
Frequency
|
Applications
Manufacturing
The 1062A Environment Monitor will ensure that your ATE system's
operating environment is properly controlled. The 1062A continuously
monitors temperature, humidity, barometric pressure, line voltage,
and line frequency. Up to 12 independent sensors plus the optional
barometric pressure sensor can be connected to the 1062A. The
sensors can be placed hundreds, even thousands, of feet away
from the 1062A. The 1062A's compact half rack size, 3.5 inches
high and 8.25 inches wide, makes it easy to rack mount directly
into your ATE system rack thus freeing up table top and bench
space.
 |
Calibration
Labs
The 1062A Environment Monitor and its sensors are traceable
to the U.S. National Institute of Standards and Technology.
Providing both high accuracy measurements and full GPIB and
RS-232-C programmability, the 1062A eliminates the need for
chart recording and maintaining cumbersome paper records.
Environmental data can be transferred directly to your calibration
systems reports and Statistical Process Control programs.
The 1062A provides you with the ability to correlate test
failures or measurement perturbations with changes in environmental
conditions, all in a "Paperless Calibration" environment.
Computer,
Control and Clean Rooms
The large display and full function keyboard make the 1062A
easy to read from a distance and easy to operate. Continuously
monitoring the operating environment of sensitive electronic
equipment, the 1062A will display the time at which the upper
and lower extremes were measured for each channel. The 1062A
supports the other 4-20 mA loop-powered or self-powered sensors
for measuring many different parameters such as pH, conductivity,
flow, position, level, current, dc voltage, tension, pneumatic
and differential pressure, RPMs, and temperature (both RTDs
and Thermocouples). Each channel's display can be customized
for your specific application.
Description
14
Channels with Flexible Displays
The 1062A offers 12 independently selectable 4-20 mA input
channels, plus an AC line voltage/frequency monitoring channel
and a dedicated channel for the optional Barometric Pressure
sensor. Each channel's parameters can be independently set
including: display title, display units, number of digits
displayed, input and output scaling factors, alarm limits,
and alarms activated. For your convenience, all parameters
and settings are preset at the factory in both English and
metric units.
Limits,
Alarms, Min.Max Readings
Every 200 milliseconds all 14 channels are individually sampled,
measured, converted to display units and compared against
selectable alarm conditions and previous minimum and maximum
readings. Any or all of the alarm indications can be activated
including: audible alert, visual display, rear panel contact
closure (or contact open), and computer SRQ interrupt over
the GPIB interface.
Built-in
Real Time Clock
The built-in real time clock provides you both current time
and time-of-event information. The time-of-event data for
each channel's minimum and maximum reading is stored in memory
for later recall.
Calibration,
Diagnostics and Special Functions
The 1062A is simple to calibrate. Only one manual adjustment
is ever required - adjusting the time base frequency. All
other adjustments can be made from the front panel keyboard
or remotely over the GPIB or RS-232-C. To insure proper operation
the 1062A performs an extensive diagnostic check of its internal
circuitry when first turned on. The 1062A offers many special
functions including: number of measurements averaged per reading,
RS-232-C setup, audible alarm sound level, internal battery
voltage, and metric or English default units.
Rear-Panel
Recorder Output
The 1062A's user selectable recorder output lets you perform
external signal processing of any of the 14 input channels.
The recorder output's scaling factor is individually settable
for each channel.
Features/Controls
Display
16 character alphanumeric vacuum-fluorescent display (VFD).
Mode
Selection
Channel 1 to 12, time, line voltage, line frequency, barometric
pressure, recorder and setup.
Setup
Parameters
Upper limit (20 mA), lower limit (4 mA), alarm parameters
(upper limit, lower limit, on/off for both limits and error
condition), title (up to 10 ASCII characters), number of decimal
places displayed to the right of the decimal point (0, 1,
or 2), mA/V "raw" display, recorder upper/lower limits, special
measurement modes, and min/max values.
Setup
Presets
English and metric units for default parameters. Preset settings
for each channel: option 01/03 temperature, option 01 humidity,
or undefined.
Min-Max
Function
Tracks minimum and maximum values and time-of-event for each
channel 1 to12, plus line voltage, frequency, and barometric
pressure. Min-max is reset using "START" function.
Real-Time
Clock
Quartz-crystal real-time clock with battery backup. Time may
be set from the front panel or via remote programming interface.
Alarm
Activated by error conditions (open, overload, etc.) on any
channel. Also activated by out-of-limits conditions on any
channel. Alarm limits may be different than the channel upper/lower
limits. All alarm conditions may be selectively disabled.
Selectable alarm indications include: visual display, audible
alert (with variable volume adjustment), contact closure (NO
or NC), and computer SRQ interrupt.
Programmability
GPIB (IEEE-488.2) and RS-232C (300 to 19200 Baud).
Chart
Recorder Output
Plot any channel's instantaneous value, scaled between limits
of your choice, on any standard chart recorder. Output is
¡¾5 volts dc.
1062A
Specifications
Inputs:
4-20
mA Inputs:
| Accuracy |
¡¾0.1%
of full scale |
| Resolution |
12
bits (0 to 25 mA) (A/D converter) |
| Supply
voltage |
(loop-powered
mode): 24 Vdc ¡¾25%;
(self-powered mode): 0 Vdc; internally-switched |
| Load
resistance |
200
Ohm + 1 diode drop, while measuring that channel
10 Ohm maximum, standby |
| Fusing |
250
mA, each input
250 mA, +24 V supply, for each group of 4 channels
Fuses are internal |
| Input
connectors |
Phoenix
pluggable type, 5 mm centers (mating connectors provided) |
| Compatible
sensors |
Temperature
Humidity
Barometric Pressure
Line Voltage
Line Frequency
pH
Conductivity
Flow
Position
Level
Current
Voltage
Tension (Strain Gauges)
Pneumatic Pressure
RPMs (Tachometer)
Specialized Temperature Ranges (RTDs & Thermocouples) |
Line
Monitor Input:
| Range |
Voltage:
0 to 400 Vrms
Frequency: 20 to 500 Hz |
| Accuracy |
Voltage:
¡¾(0.25% + 1 Vrms), at 47 to 63 Hz
Frequency: ¡¾0.027% |
| Resolution |
Voltage:
0.1 Vrms, maximum
Frequency: 0.01 Hz, maximum |
| Input
impedance |
2
Mohm, balanced to ground |
| Input
connectors |
Standard
banana jacks |
Outputs:
Alarm
Outputs:
| Relay |
NO/NC
(SPDT) |
| Rating |
24
Vac or Vdc, 250 mA maximum |
| Connectors |
Phoenix
pluggable type, 5 mm centers (mating connectors provided) |
Recorder
Outputs:
| Voltage |
¡¾5
Vdc |
| Load |
200
Ohm minimum |
| Connectors |
Phoenix
pluggable type, 5 mm centers
(mating connectors provided) |
General:
Physical:
| Supply
Power |
100/200/220/240
Vac ¡¾10%
47 to 63 Hz single phase
30 VA maximum |
| Weight |
4
kg (9 lbs), net
7 kg (16 lbs), shipping |
| Dimensions |
100
mm H x 210 mm W x 400 mm D (3.95 in. x 8.25 in. x 15.8
in.) (with feet and handles) |
Communications
| IEEE-488 |
SH1,
AH1, T6, L4, SR1, RL1, PP0, DC1, DT0, C0, E2 |
Environmental:
| Temperature
range |
0¡Æ
C to 50¡Æ C, operating
-40¡Æ C to +75¡Æ C, non-operating |
Calibration:
| Interval |
One
(1) year |
| Traceability |
The
1062A Environment Monitor's measurement functions are
traceable to the U.S. National Institute of Standards
and Technology. We provide a Certificate of Compliance
and Traceability evidencing traceability at the time
of purchase and whenever maintenance or calibration
are performed at our facility. |
Option
Specifications:
Specifications
for these options are derived from the instrumentation uncertainties
of
the Model 1062A, and the manufacturer's specifications for
the sensors. Performance verification at Arbiter Systems,
Inc. is limited to the measurement of the 1062A instrumentation
uncertainties.
Option
01 RH/Temperature Sensor:
| Manufacturer's
part no. |
3HMW20YB,
Vaisala Sensor Systems Div.; Helsinki Finland; Woburn,
MA USA |
| Relative
Humidity |
| Range |
0
to 100% RH |
| Accuracy |
¡¾2.1%
RH, 0 to 90% RH
¡¾3.1% RH, 90 to 100% RH
(at 20¡Æ C, including 1062A instrumentation uncertainty) |
| Resolution |
¡Â
0.04% RH |
| Temperature
coefficient |
¡¾0.04%
RH/¡Æ C |
| Response
time |
15
seconds |
| Temperature |
| Range |
-5¡Æ
C to +55¡Æ C |
| Accuracy |
¡¾0.26¡Æ
C (at 20¡Æ C, including 1062A instrumentation uncertainty) |
| Resolution |
¡Â
0.03¡Æ C |
| Temperature
coefficient |
¡¾0.02¡Æ
C/¡Æ C |
| Linearity |
<
0.2¡Æ C |
| Sensor |
RTD,
Pt 100 Ohm 1/3 DIN 43760B |
| Housing |
| Size |
110
mm x 85 mm x 32 mm (4.33 in. x 3.35 in. x 1.26 in.)
nominal |
| Material |
ABS
thermoplastic |
| Mounting |
Wall
mount |
| Weight |
115
g (4 oz) nominal |
| Cable |
| Specification |
30
m (100 ft) Belden 8443 o/e plus mating connectors |
Option 02 Barometric Pressure Sensor:
-
| Manufacturer's
part no. |
PTA427,
Vaisala Sensor Systems Div.; Helsinki Finland; Woburn,
MA USA |
| Range |
800
to 1060 hPa (800 to 1060 mbar) |
| Total
accuracy |
0.6
hPa (at 20¡Æ C, including 1062A instrumentation uncertainty
and drift for one year) |
| Temperature
coefficient |
¡¾0.002
hPa/¡Æ C at 1000 hPa
¡¾0.0001 hPa/hPa * ¡Æ C |
| Stability |
0.3
hPa/year |
| Stability
due to shock |
¡¾0.3
hPa |
| Resolution |
¡Â
0.06 hPa |
| Underpressure
limit |
500
hPa (5500 m or 18,000 ft altitude) |
-
-
- Option
03 Temperature Sensor:
-
| Manufacturer's
part no. |
TT155PD1GW/S472PDY4K0,
Minco Products Inc., Minneapolis, MN USA |
| Accuracy |
¡¾0.51¡Æ
C (including 1062A instrumentation uncertainty and
temperature -5 to +55 ¡Æ C) |
| Resolution |
¡Â
0.04¡Æ C |
| Sensor |
RTD,
Pt 100 Ohm, per DIN 43760, class B |
| Housing |
| Size |
115
mm x 70 mm x 40 mm (4.5 in. x 2.75 in. x 1.56 in.)
nominal |
| Material |
High-impact
thermoplastic with aluminum faceplate |
| Mounting |
Wall
mount |
| Cable |
30
m (100 ft) Belden 8442 o/e plus mating connectors |
-
-
-
- Rack
Mount Kits:
-
| 11001A |
Rack
Mount Kit, single unit |
| 11002A |
Rack
Mount Kit, two units side-by-side |
 |
|